Sterilisers
Medicine has the goal of protecting and saving lives. Therefore, disinfection of receptacles and instruments is not enough. The setpoint-dependent programme resume function SetpointWAIT of Memmert hot air sterilisers guarantees precise sterilisation times and the complete killing off of even the most resistant microorganisms. The appliances comply with all relevant national and international standards and requirements for medical devices and can be validated without problems.
- Adjustment of pre-heated fresh air admixture by air flap control in 10 % steps.
- Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display.
- Self-diagnostics for fault analysis.
- One Pt100 sensor DIN class A.
- Digital backwards counter with target time setting, adjustable from 1 minute to 99 days.
- Parameters adjustable on ControlCOCKPIT: temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime.
- SetpointWAIT function - the process time does not start until the set temperature is reached.
- Resolution of display for setpoint values 0.1 °C up to 99.9 °C, 0.5 °C from 100 °C and for actual values 0.1 °C (LED).
- Programme stored in case of power failure.
- Delivered with works calibration certificate for +160 °C.
- Adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20 °C above nominal temperature.
- Fully insulated stainless steel door with 2-point locking.
- Easy-to-clean interior, made of stainless steel with large-area heating of four sides.
Sterilisers configurations:
- SN models - sterilisers with natural convection.
- SF models - sterilisers with forced air circulation.
Sterilisers control variants:
- Standard SingleDISPLAY models is recommended for standard temperature control applications with insensitive load.
- plus TwinDISPLAY models are recommended for maximum convenience with complex processes and programming features.
Model/Parameter | SN30 SF30 SN30plus SF30plus | SN55 SF55 SN55plus SF55plus | SN75 SF75 SN75plus SF75plus | SN110 SF110 SN110plus SF110plus | SN160 SF160 SN160plus SF160plus | SN260 SF260 SN260plus SF260plus | SN450 SF450 SN450plus SF450plus | SN750 SF750 SN750plus SF750plus | |
Volume, l | 32 | 53 | 74 | 108 | 161 | 256 | 449 | 749 | |
Interior width, mm | 400 | 400 | 400 | 560 | 560 | 640 | 1040 | 1040 | |
Interior height, mm | 320 | 400 | 560 | 480 | 720 | 800 | 720 | 1200 | |
Interior depth, mm | 250 | 330 | 330 | 400 | 400 | 500 | 600 | 600 | |
Max. number of shelves | 3 | 4 | 6 | 5 | 8 | 9 | 8 | 14 | |
Max. loading per shelf, kg | 20 | 20 | 20 | 20 | 20 | 20 | 30 | 30 | |
Max. loading of chamber, kg | 60 | 80 | 120 | 175 | 210 | 300 | 300 | 300 | |
Exterior width, mm | 585 | 585 | 585 | 745 | 745 | 824 | 1224 | 1224 | |
Exterior height, mm | 704 | 784 | 944 | 864 | 1104 | 1183 | 1247 | 1720 | |
Exterior depth, mm | 434 | 514 | 514 | 584 | 584 | 684 | 784 | 784 | |
Electrical load at 230 V, 50/60 Hz, W | 1600 | 2000 | 2500 | 2800 | 3200 | 3400 | |||
Electrical load at 400 V, 50/60 Hz, W | 5800 | 7000 | |||||||
Net weight, kg | 46 | 57 | 66 | 74 | 96 | 110 | 161 | 217 |