Analüütilised kaalud ABT
- Automatic internal adjustment in the case of a change in temperature > 0,5 °C and time-controlled every 4 hours, guarantees high degree of accuracy and makes the balance independent of its location of use
- Dosage aid: High-stability mode and other filter settings can be selected
- Simple recipe weighing and documenting with a combined tare/print function. In addition, the ingredients for the recipe are numbered automatically and printed out with their corresponding number and nominal weight
- Identification number: 4 digits, printed on calibration protocol freely programmable
- Automatic data output to the PC/printer each time the balance is steady
- Large glass draught shield with 3 sliding doors for easy access to the items being weighed
Single-cell advanced technology:
- Fully automatic manufactured weighing cell from one piece of material
- Stable temperature behaviour
- Short stabilisation time: Steady weight values within approx. 4 sec under laboratory conditions (on all models with readout [d] = 0,1 mg, 10 s on all models with readout [d] = 0,01 mg)
- Shock proof construction
- High corner load performance[/]Protective working cover included with delivery
Model | Weighing capacity g | Read- ability mg | Verific. value mg | Minimum load g | Repeatability
mg | Linearity
mg | Option: Verification | Option: DAkkS Calibration certificate |
ABT 120-4NM | 120 | 0,1 | 1 | 10 | 0,1 | 0,2 | optional | optional |
ABT 220-4NM | 220 | 0,1 | 1 | 10 | 0,1 | 0,2 | optional | optional |
ABT 320-4NM | 320 | 0,1 | 1 | 10 | 0,1 | 0,3 | optional | optional |
ABT 100-5NM | 101 | 0,01 | 1 | 1 | 0,05 | 0,15 | optional | optional |
ABT 120-5DNM | 42 | 120 | 0,01 | 0,1 | 1 | 1 | 1 | 0,02 | 0,1 | 0,05 | 0,2 | optional | optional |
ABT 220-5DNM | 82 | 220 | 0,01 | 0,1 | 1 | 1 | 1 | 0,05 | 0,1 | 0,1 | 0,2 | optional | optional |
Note: For applications that require verification, please order verification at the same time, initial verification at a later date is not possible. Verification at the factory, we need to know the full address of the location of use.
DNM models - Dual-range balance (D=Dual): switches from readout d1 to readout d2 automatically.